Boonton 4500B RF Peak Power Analyzer
Operation
4-75
Table 4-22. Util > Remote Conf > Emulation > Submenu
Menu Item
Function
(Type)
Selections
(SCPI CMD)
Boot Language
SCPI, 4500A
Boot Language selects the default language
(Toggle)
at instrument power-up. It does not change
the Current Language.
Curr Language
SCPI, 4500A
Current Language permits viewing and
(Toggle)
changing the language at any time without a
remote command. At power-up, the Current
Language is set to the Boot Language.
SYSTem:LANGuage EMUL4500A
*IDN Response
4400, 4400A
Selects model number the instrument returns
(Mult. Choice)
4500, 4500A
in Emulation mode to *IDN? query.
4500B
Option Select
None, -S/1,
Selects the option identifier appended to
(Mult. Choice)
-04, -05, 06
the Emulation mode *IDN? query.
Ignore Unsupp
On, Off
In 4400/4500/4400A/4500AEmulation mode
(Toggle)
only, if ―On‖, ignores GPIB bus when an
unsupported command is received.
If ―Off‖, returns error message.
Fig. 4-38. Util >
Remote Conf >
Emulation > submenu
Note
For compatibility purposes, the 4500B remote interface can be set to emulate the BOONTON 4400, 4500, 4400A
(including 4400A-L) and 4500A (including 4500A-L) series of instruments. For additional discussion of this
subject, see Section 5.2. To use the 4500B SCPI interface, set the Util > Remote Conf > Emulation > Curr
Language menu selection to SCPI. To emulate the 4400/4500 legacy interface, set the Util > Remote Conf >
Emulation > Curr Language menu selection to the desired model. To control the language in use at instrument
power-up, select your choice in the in the Util > Remote Conf > Emulation > Boot Language menu box.