Table 4-34. calibrator > extensions > submenu, Table 4-34. calibrator > extensions > submenu -87, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual
Page 159: Operation
Boonton 4500B RF Peak Power Analyzer
Operation
4-87
Figure 4-48.
Table 4-34. Calibrator > Extensions > Submenu
Menu Item
Function
(Type)
Selections
(SCPI Cmd)
Level Step
0.1 to 60 dB
Selects the step level for the calibrator.
(Numeric)
The step level value is incremented or decremented from the current
calibrator when the knob or arrow keys are activated.
OUTPut:LEVel:STEP <0.1 to 60.0>
Calibrator
REPORT
Includes information about the installed calibrator.
Press the Calibrator > Extensions > Calibrator menu key to list
calibrator information. The report includes: Status, Serial Number,
Software Version, Calibration Date and Internal Temperature.
A non-zero status reading indicates that an error condition exits.