Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual
Page 108
Boonton 4500B RF Peak Power Analyzer
Operation
4-36
Table 4-7D. Ext Trig # > Extensions > Submenu (continued)
Menu Item
Function
(Type)
Selections
(SCPI cmd)
Define Pulse
Pulse & Mod
Selects pulse level criteria for pulse time measurements
(Submenu)
MENU
Identical to Chan # > Exten > Define Pulse menu. See Table 4-4.
Video BW
Pulse, Mod & Stat
Selects either the narrowband or wideband external trigger response,
(Toggle)
Low, High
Press Ext Trig # > Extensions > Video BW > Low to select a low
bandwidth trigger response that is useful for triggering on or displaying
noisy, low frequency signals..
Press Ext Trig # > Extensions > Video BW > High to select a wideband
response that is useful for triggering on or displaying high frequency
signals, clean signals of any bandwidth or short pulses.
SENSe[3|4]:BANDwidth {LOW, HIGH}
Averaging
Pulse
Selects the number of samples that are averaged at each point
(Numeric)
1 to 16,384
Press Ext Trig # > Extensions > Averaging to activate this function.
Use any of the data entry controls to select the number of samples to be
averaged at each point of the waveform to produce the waveform
display. Default = 8.
SENSe[3|4]:AVERage <1 to 16384>