1 sense:caltab data array commands, Sense:caltab:count, Sense:caltab:data[:next – Boonton 4500B Peak Power Meter User Manual
Page 267: Sense:caltab:index, Sense:correction:calfactor, 1 sense:caltab data array commands -67, Sense:correction:calfactor -67, Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/267/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-67
5.5.14.1 SENSe:CALTAB Data Array Commands
The purpose of the SENS:CALTAB group of commands is to return the power calibration table (or ―caltable‖), which is the
same size as the histogram (16,384 points), and contains the power level for each corresponding bin in the histogram. The
power values are not equally spaced, so it is necessary to know both the histogram and caltable arrays to compute statistical
information about the input signal.
SENSe:CALTAB:COUNt
Description:
Set or return the number of cal table values to be returned as data by SENSe:CALTAB:DATA?
Syntax:
SENSe[1|2]:CALTAB:COUNt
Argument:
Valid Modes:
All
SENSe:CALTAB:DATA[:NEXT]?
Description:
Return ―count‖ power values from the specified channel‘s cal table (if any) starting at ―index‖.
Units are milliwatts without regard to the channel‘s units setting.
Syntax:
SENSe[1|2]:CALTAB:DATA?
Argument:
None, query only.
Valid Modes:
All
SENSe:CALTAB:INDEX
Description:
Set or return the index of the first cal table value to be returned as data by
SENSe:CALTAB:DATA?
Syntax:
SENSe[1|2]:CALTAB:INDEX
Argument:
Valid Modes:
All
SENSe:CORRection:CALFactor
Description:
Set or return the frequency calfactor currently in use on the selected channel. Note setting a
calfactor with this command will override the ―automatic‖ frequency calfactor that was calculated
and applied when the operating frequency was set, and setting the operating frequency will
override this calfactor setting.
Syntax:
SENSe[1|2]:CORRection:CALFactor
Argument:
Valid Modes:
All