Read:array:marker:power, Read:array:marker:window, Read:cw:power – Boonton 4500B Peak Power Meter User Manual
Page 260: Read:interval:average, Read:array:marker:power? -60, Read:array:marker:window? -60, Read:cw:power? -60, Read:interval:average? -60, Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/260/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-60
READ:ARRay:MARKer:POWer?
Description:
Returns an array of the current marker measurements for the specified channel. The array consists
of the average, maximum, and minimum power and peak-to-average ratio between the two
markers, powers at both markers, and the ratio of the two markers. Note the peak-to-average ratio
and marker ratio are returned in dB for log units, and percent for linear units.
Syntax:
READ[1|2|3|4|6|7]:ARRay:MARKer:POWer?
Returns:
CC1, Pavg, CC2, Pmax, CC3, Pmin, CC4, PkToAvgRatio, CC5, Pwr@Marker1, CC6,
Pwr@Marker2, CC7, Mrk1/Mrk2 ratio
Where the CCn‘s are the measurement condition codes for each measurement.
Valid Modes:
Pulse and Modulated
READ:ARRay:MARKer:WINDow?
Description:
Return the current marker window display contents, MK1, MK2 and marker math result. The
channel assignment(s), marker math function and basic units are the same as the display.
Syntax:
READ:ARRay:MARKer:WINDow?
Returns:
CC, MK1, MK2 and marker math display
Where CC is the measurement condition code.
Valid Modes:
All
READ:CW:POWer?
Description:
Return current average amplitude reading in channel units.
Syntax:
READ[1|2|3|4|5|6|7]:CW:POWer?
Returns:
CC, Average power (watts, dBm) or average voltage (volts, dBv)
Where CC is the measurement condition code.
Valid Modes:
Modulated mode only.
READ:INTERval:AVERage?
Description:
For the specified channel, return the average power or voltage between marker 1 and marker 2.
The units will be the same as the specified channel.
Syntax:
READ[1|2|3|4|5|6|7]:INTERval:AVERage?
Returns:
CC, average power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated