Table 4-17. meas > menu (statistical), Table 4-17. meas > menu (statistical) -67, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual
Page 139: Operation
![background image](/manuals/401804/139/background.png)
Boonton 4500B RF Peak Power Analyzer
Operation
4-67
Table 4-17. Meas > Menu (Statistical)
Menu Item
Function
(Type)
Selections
(SCPI cmd)
Freq Track
Stat
Selects the channel frequency entry mode.
(Toggle)
On, Off
When frequency tracking is ‖On‖, the same
frequency is assigned to both channels.
Changing either CH1 or CH2 frequencies
changes both of them. Both windows are
updated.
When frequency tracking is ―Off‖, the
frequency for each channel can be
individually assigned.
SENSe:CORRection:FRQTRK {ON, OFF}
Freq CH1
Stat
Selects the input signal frequency on CH1
(Numeric)
Continuous Range:
for frequency response correction.
(Sensor Dependent)
See Table 4-15 above.
Freq CH2
Stat
Selects the input signal frequency on CH2
(Numeric)
Continuous Range:
for frequency response correction.
(Sensor Dependent)
See Table 4- 15 above.
Sys Tracking
Stat
Controls the tracking of channel parameters.
(Toggle)
Off, On
When ―On‖, Channel 1 and Channel 2 units
and video bandwidth track any changes made
on either channel.
Fig. 4-33a. Meas >
CALCulate:SYSTRK {ON, OFF}
menu
Normalize Pwr
Stat
Select display and marker scaling of distributions
(Toggle)
Off, On
When ―Off‖, power distributions are displayed as power versus
probability. When ―On‖, power distributions are displayed as
normalized power versus probability. Each power point is divided by
the average power of all the points. Marker measurements also
change from power (dBm or watts) at a point to normalized power at
that point (dBr or percent).
SENSe:NORMalize {ON, OFF }