Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual
Page 325
Boonton 4500B RF Peak Power Analyzer
Application Notes
6-15
The following two global status values are displayed:
Text
TERM
DEFINITION
Total
Time
The total time in Hours:Minutes:Seconds that the data acquisition has been
running.
Samples
The total number of data samples that has been acquired for each channel in the
current run.
Note
The sampling speed of Channel 1 and Channel 2 is 25 MSa/sec. The sampling speed of the external
Trigger 1
and Trigger 2 inputs is 2 MSa/sec. If Channel 1 and Channel 2 are running, the sampling speed of each is
still 25
MSa/sec. If Channel 1 and Trigger 1 are running, the sampling speed for both is 2 MSa/sec. The same is
true for
Channel 2 and Trigger 2. For fastest sampling speed, do not run the CDF (or 1-CDF) of a channel and its
external
trigger at the same time.