5.5.11 OUTPut Subsystem
The OUTPut group of commands is used to control the internal 1GHz RF calibration source. The internal calibrator can be
configured as a programmable pulse source, with options for modulation rate and duty cycle. It can also be driven by an
external pulse modulation source.
OUTPut:LEVel:POWer
Description:
Set or return the power level for the 1 GHz calibrator output signal.
Syntax:
OUTPut:LEVel:POWer
Argument:
Valid Modes:
Any
OUTPut:LEVel:LIMit
Description:
Set or return the limit power level in dBm for the 1 GHz calibrator output signal. The calibrator
output level cannot be set to exceed this limit.
Syntax:
OUTPut:LEVel:LIMit
Argument:
Valid Modes:
Any
OUTPut:LEVel:STEP
Description:
Set or return the power step size in dB used by the Knob and Arrow keys on the front panel to
control the output level of the 1 GHz calibrator.
Syntax:
OUTPut:LEVel:STEP
Argument:
Valid Modes:
Any