MARKer:POSItion:PIXel
Description:
Set or return the horizontal pixel position (X-axis-position) of the selected vertical marker. There
are 501 pixel positions numbered from 0 to 500 inclusive.
Syntax:
MARKer[1|2]:POSItion:PIXel
Argument:
Valid Modes:
All
MARKer:POSItion:TIMe
Description:
Set or return the time (X-axis-position) of the selected vertical marker relative to the trigger. Note
that time markers must be positioned within the time limits of the trace window in the graph
display. If a time outside of the display limits is entered, the marker will be placed at the first or
last time position as appropriate.
Syntax:
MARKer[1|2]:POSItion:TIMe
Argument:
Valid Modes:
Pulse and Modulated modes.
Restrictions:
For zero delay trigger position in the center of the display , the following relationship must be
satisfied: TrigDly - (5*time/div) < MarkerTime < TrigDly + (5*time/div) where the timebase