Page 226: Remote operation
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-26
DISPlay:PARAmeter:TOP:RIGHt
Description:
Set or return the auto-measure or fixed parameter displayed in the Top row, Right column of the
nine element parameter field above the trace box. The selected parameter will be for the currently
active channel. See DISPlay:ACTive command. DISPlay:PARAmeter:MODE must be set to
MEAS.
Syntax:
DISPlay:PARAmeter:TOP:RIGHt
Argument:
See Display Parameter Table below
Valid Modes:
Pulse Mode only. Note: Modulated mode does not display Status and Measurement parameters.
Statistical mode displays a fixed combination of status and auto-measure parameters.
DISPLAY PARAMETER TABLE
function {
PWIDTH
Pulse width
RISETIME
Risetime
FALLTIME
Falltime
PERIOD
Period *
PRFREQ
Pulse repetition frequency *
DUTY
Duty cycle *
OFFTIME
Pulse offtime *
PEAK
Peak Power
PULSE
Pulse Power *
OVRSHT
Pulse overshoot
AVGPWR
Average power
TOPPWR
Pulse top level power
BOTPWR
Pulse bottom level power
SKEW
Trigger skew
EDGEDLY
Edge delay
FREQ
Measurement frequency setting
VIDBW
Sensor video bandwidth setting
AVG
Channel averaging value
TRA
A Trigger slope, source and level
TRB
B Trigger slope, source and level or off
TRDLY
Trigger delay setting
VSCALE
Vertical scale setting
CENTER
Vertical center setting
OFFSET }
Vertical offset setting
* at least one complete waveform cycle must be displayed