Marker:math:each, Marker:position:percent, Marker:math:each -46 – Boonton 4500B Peak Power Meter User Manual
Page 246: Marker:position:percent -46, Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/246/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-46
MARKer:MATH:EACH
Description:
Set the marker group to EACH and configure the marker math function.
Syntax:
MARKer:MATH:EACH
Argument:
Specifies the result displayed in the center marker window
RATIO -the ratio of the left and right marker window readings in dBr or percent.
DELTA -valid only with
and right marker window readings expressed in channel units
Specifies how the left and right marker windows are used to calculate the center marker
window.
MK1MK2 – ratio is MK1/MK2; difference is MK1-MK2
MK2MK1 - ratio is MK2/MK1; difference is MK2-MK1
{dBm, W, V, dBV, dBmV, dBuV}
Valid Modes:
Modulated, Pulse and Statistical (RATIO with MK1MK2 only) modes
NOTE
There is no query form for MARKer:MATH:BOTH or MARKer:MATH;EACH.
MARKer:POSItion:PERcent
Description:
Set or return the percent probability (X-axis-position) of the selected vertical marker. Note that
the power value returned for each marker will depend on the setting of CALCulate:MODE.
When set to CDF, the highest power levels are towards the right side of the screen, with maximum
(highest peak) power occurring at 100%. When set to CCDF (also called 1-CDF), the highest
levels are towards the left, with peak power at 0%.
Syntax:
MARKer[1|2]:POSItion:PERcent
Argument:
Valid Modes:
Statistical modes only.