Boonton 4500b rf peak power analyzer, Maintenance – Boonton 4500B Peak Power Meter User Manual
Page 348
Boonton 4500B RF Peak Power Analyzer
Maintenance
7-16
Table 7-3a. Calibrator Linearity - High Power Range
Cal Level Minimum Measured Maximum
(dBm) (dBm) (dBm) (dBm)
20.0 19.84 __________ 20.16
15.0 14.84 __________ 15.16
10.0 9.84 __________ 10.16
Table 7-3b. Calibrator Linearity - Low Power Range
Cal Level Minimum Measured Maximum
(dBm) (dBm) (dBm) (dBm)
5.0 4.9 __________ 5.1
-5.0 -5.1 __________ -4.9
-10.0 -10.16 __________ -9.84
-15.0 -15.16 __________ -14.84
-20.0 -20.25 __________ -19.75
-25.0 -25.25 __________ -24.75
-30.0 -30.25 __________ -29.75