Fetch:array:temperature:caltemp, Fetch:array:temperature:current, Fetch:cw:power – Boonton 4500B Peak Power Meter User Manual
Page 236: Fetch:interval:average, Fetch:array:temperature:caltemp? -36, Fetch:array:temperature:current? -36, Fetch:cw:power? -36, Fetch:interval:average? -36, Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/236/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-36
FETCh:ARRay:TEMPerature:CALTemp?
Description:
Return the internal temperature of the Channel 1 and Channel 2 peak power sensors at the at the
time of autocalibration. Note: If a non-temperature compensated peak sensor is in use, the 4500B
displays a warning message if the temperature has drifted more than 4C from the autocal
temperature.
Syntax:
FETCh:ARRay:TEMPerature:CALTemp?
Returns:
CC1, Sensor1 AutoCal Temp in degrees C, CC2, Sensor2 AutoCal Temp in degrees C
Where CC1 and CC2 are the measurement condition codes.
Valid Modes:
All
FETCh:ARRay:TEMPerature:CURRent?
Description:
Return the current internal temperature of the Channel 1 and Channel 2 peak power sensors. Note:
If a non-temperature compensated peak sensor is in use, the 4500B displays a warning message if
the temperature has drifted more than 4C from the autocal temperature.
Syntax:
FETCh:ARRay:TEMPerature:CURRent?
Returns:
CC1, Sensor1 Temp in degrees C, CC2, Sensor2 Temp in degrees C
Where CCn is the measurement condition code.
Valid Modes:
All
FETCh:CW:POWer?
Description:
Return current average amplitude reading in channel units.
Syntax:
FETCh[1|2|3|4|5|6|7]:CW:POWer?
Returns:
CC, average power (watts, dBm) or average voltage (volts, dBv)
Where CC is the measurement condition code.
Valid Modes:
Modulated and Statistical modes.
FETCh:INTERval:AVERage?
Description:
For the specified channel, return the average power or voltage between marker 1 and marker 2.
The units will be the same as the specified channel.
Syntax:
FETCh[1|2|3|4|5|6|7]:INTERval:AVERage?
Returns:
CC, average power or voltage between markers
Where CC is the measurement condition code.
Valid Modes:
Pulse and Modulated