Trg -10, Tst? -10, Wai -10 – Boonton 4500B Peak Power Meter User Manual
Page 210: Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/210/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-10
*TRG
Description:
Simulate the bus trigger command.. This command has the same effect as GET, Group Execute
Trigger, except it must be parsed and decoded before the action takes place. There is no query
form of this command.
Syntax:
*TRG
Argument:
None
Valid Modes:
Any
*TST?
Description:
Self-test query. This command initiates a self-test of the instrument, and returns a result code
when complete. The result is zero for no errors, or a signed, 16-bit number if any errors are
detected.
Syntax:
*TST
Returns:
Error Code
Valid Modes:
Any
*WAI
Description:
Wait command. This command insures sequential, non-overlapped execution. The 4500B always
operates in non-overlapped, sequential mode. The command is accepted as valid but takes no
action.
Syntax:
*WAI
Argument:
None
Valid Modes:
Any