Table 4-9a. trig > b trig > submenu, Table 4-9a. trig > b trig > submenu -47, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual
Page 119: Operation
![background image](/manuals/401804/119/background.png)
Boonton 4500B RF Peak Power Analyzer
Operation
4-47
Table 4-9. Trig > Menu (continued)
Menu Item
Function
(Type)
Selections
(SCPI cmd)
To solve this problem, the Trig > HoldOff
function specifies a time period during which triggering is inhibited. By
setting the Trig > HoldOff period to be slightly less than the frame
interval, the instrument can be made to ―lock on‖ to the framing pattern
and present a stable display. The minimum HoldOff value is 100 ns, the
minimum interval after a trigger event before the next trigger can
occur. Entering ―0‖ as the HoldOff parameter disables HoldOff.
TRIGger:HOLDoff <0 to 65000e-6> (sec)
B trigger
Pulse & Stat*
Displays a menu for the B Trigger
(Submenu)
MENU
Table 4-9A. Trig > B Trig > Submenu
Menu Item
Function
(Type)
Selections
(SCPI cmd)
B Mode
Pulse & Stat*
Selects the operating mode of the B Trigger
(Mult. Choice)
Off,
B Dly by Time
The B trigger or delayed trigger provides a means to delay the start of
B Dly by Evnt
sweep, after the receipt of the A trigger or delaying trigger, for a
specified time period or after a specified number of trigger events. The
A and B triggers may have the same or different signal source, level
and slope.
B Dly by Time – This mode delays the start of sweep, after the A
trigger, by the time specified in Trig > B Trig > B Dly Time.
B Dly by Evnt – This mode delays the start of sweep, after the A
trigger, by the number of B trigger events specified in
Trig > B Trig > B Dly EvntCnt.
Select Off to disable the B trigger.
TRIGger:BTRIGger:MODE {OFF, TDLY, EDLY}