Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-89
TRACe:SEND
Description:
Send the trace data for the source selected by TRACe:SOURce to the file specified by the
argument string. The string must contain the entire path, a filename without an extension, and be
enclosed in quotes. The application will append the .TXT extension automatically. The data
delimiter is set by TRAce:DELImiter. If a folder specified in the path does not exist, the error
―Folder Not Found‖ will be set. If the file already exists, it can only be overwritten if the
command MEMory:OVERwrite ON has been issued; otherwise, the ―File Already Exists‖ error
message will be set.
Syntax:
TRACe:SEND
Argument:
Note: The internal hard drive is always C: The operating system will assign D:, E:, etc. to USB
drives and other external drives.
Valid Modes:
All
TRACe:SOURce
Description:
Set or return the source of a trace array or histogram array (Statistical mode only).
Syntax:
TRACe:SOURce
Argument:
Valid Modes: