Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual
Page 184
Boonton 4500B RF Peak Power Analyzer
Operation
4-112
Table 4-53. Spcl > Servicing > Time Skew Adj > Submenu
Menu Item
Function
(Type)
Selections
(SCPI cmd)
CH1 Skew Adj
0.0 to ± 99.9 nsec
Adjust the time skew of channel 1
(Numeric)
Press the Spcl > Servicing > Time Skew > CH1 Skew Adj menu key to
display a menu to adjust the timing skew between channel 1 and the
trigger system time reference.
CH2 Skew Adj
0.0 to ± 99.9 nsec
Adjust the time skew of channel 1
(Numeric)
Press the Spcl > Servicing > Time Skew > CH2 Skew Adj menu key to
display a menu to adjust the timing skew between channel 2 and the
trigger system time reference.