Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-61


READ:INTERval:MAXimum?

Description:

For the specified channel, return the maximum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.

Syntax:

READ[1|2|3|4|6|7]:INTERval:MAXimum?

Returns:

CC, maximum instantaneous power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:INTERval:MINimum?

Description:

For the specified channel, return the minimum instantaneous power or voltage between marker 1
and marker 2. The units will be the same as the specified channel.

Syntax:

READ[1|2|3|4|6|7]:INTERval:MINimum?

Returns:

CC, minimum instantaneous power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:INTERval:MAXFilt?

Description:

For the specified channel, return the maximum filtered power or voltage between marker 1 and
marker 2. The units will be the same as the specified channel.

Syntax:

READ[1|2|3|4|6|7]:INTERval:MAXFilt?

Returns:

CC, maximum filtered power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

READ:INTERval:MINFilt?

Description:

For the specified channel, return the minimum filtered power or voltage between marker 1 and
marker 2. The units will be the same as the specified channel.

Syntax:

READ[1|2|3|4|6|7]:INTERval:MINFilt?

Returns:

CC, minimum filtered power or voltage between markers

Where CC is the measurement condition code.