5-38

FETCh:INTERval:PKAVG?

Description:

For the specified channel, return the peak-to-average ratio of the power or voltage between marker
1 and marker 2. The units are dB for logarithmic channel units or percent for linear channel units.

Syntax:

FETCh[1|2|6|7]:INTERval:PKAVG?

Returns:

CC, peak-to-average ratio of power or voltage between markers

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated

FETCh:KEY?

Description:

Return the key code of the last key depressed; e.g. ESC = 8.

Syntax:

FETCh:KEY

Returns:

CC, key code

Where CC is the measurement condition code.

Valid Modes:

All


FETCh:MARKer:AVERage?

Description:

For the specified channel and marker, return the average power or voltage at the marker. The
units are the same as the specified channel.

Syntax:

FETCh[1|2|3|4|5|6|7]:MARKer[1|2]:AVERage?

Returns:

CC, average power or voltage at marker

Where CC is the measurement condition code.

Valid Modes:

Pulse and Modulated.

FETCh:MARKer:DELTa?

Description:

For the specified channel return the difference between MK1 and MK2. The units will be the same
as marker units.

Syntax:

FETCh[1|2|3|4|5|6|7]:MARKer:DELTa?

Returns:

CC, (MK1 – MK2)

Where CC is the measurement condition code.

Valid Modes:

All