Display:pulse:timebase, Display:screensaver:brightness, Display:screensaver:state – Boonton 4500B Peak Power Meter User Manual
Page 229: Display:screensaver:time, Display:pulse:timebase -29, Display:screensaver:brightness -29, Display:screensaver:state -29, Display:screensaver:time -29, Boonton 4500b rf peak power analyzer, Remote operation
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-29
DISPlay:PULSe:TIMEBASE
Description:
Set or return the pulse mode timebase in seconds/division. The 4500B has fixed timebase settings
in a 1-2-5 sequence, and if the argument does not match one of these settings, it will be forced to
the next highest entry.
Syntax:
DISPlay:PULSe:TIMEBASE
Argument:
30 sec/div,
1 to 10 min/div,
30 min/div and 60 min/div
Valid Modes:
Pulse mode and Gated Statistical mode
Note: There are separate timebases for the Pulse/Gated Statistical mode and the Modulated mode.
The arguments selected are saved and restored independently by mode.
DISPlay:SCREensaver:BRIGhtness
Description:
Set or return the display backlight brightness in percent when dimmed..
Syntax:
DISPlay:SCREensaver:BRIGhtness
Argument:
Valid Modes:
All
DISPlay:SCREensaver:STATe
Description:
Set or return the state of the screensaver function. When ON, the screensaver dims the LCD
backlight to a specified intensity after a specified period of keyboard inactivity. GPIB activity
does not disturb display backlighting unless DISPlay:SCREensaver commands are sent.
Syntax:
DISPlay:SCREensaver:STATe
Argument:
Valid Modes:
All
DISPlay:SCREensaver:TIME
Description:
Set or return the time in minutes of no activity that dims the display backlight.
Syntax:
DISPlay:SCREensaver:TIME
Argument:
Valid Modes:
All