beautypg.com

Memory:sys:store, Mmemory:copy, Mmemory:load – Boonton 4500B Peak Power Meter User Manual

Page 252: Memory:sys:store -52, Mmemory:copy -52, Mmemory:load -52, Boonton 4500b rf peak power analyzer, Remote operation

background image

Boonton 4500B RF Peak Power Analyzer

Remote Operation

5-52

MEMory:SYS:STORe

Description:

Store an instrument configuration to the filename specified by the <‖asc‖> ASCII string. The
string must contain a filename with the entire path, without the extension, enclosed in quotes.

Syntax:

MEMory:SYS:STORe <‖asc‖>

Argument:

<‖asc‖> = ―C:\prog\dat\image4500B‖ (example)

Valid Modes:

Any

MMEMory:COPY

Description:

Copy a waveform from a channel specified by the string to a reference channel specified
by the string. A space is required between the two arguments.

Syntax:

MMEMory:COPY sp

Argument:

= {CH1, CH2, TRIG1, TRIG2}

= {MEM1, MEM2}

Valid Modes:

Any

MMEMory:LOAD

Description:

Recall a stored waveform from the filename specified by the <‖asc1‖> string and load it into
reference channel MEM1 or MEM2 specified by the string. The <‖asc1‖> string must
contain a filename with the entire path, without the extension, enclosed in quotes, and a reference
channel destination. A space is required between the two arguments.

Syntax:

MMEMory:LOAD <‖asc1‖> sp

Argument:

<‖asc1‖> = ―C:\prog\dat\WFM001‖ (example)

= {MEM1, MEM2}

Valid Modes:

Any