Calculate:math:arga, Calculate:math:argb, Calculate:math:operator – Boonton 4500B Peak Power Meter User Manual
Page 212: Calculate:mode, Calculate:math:arga -12, Calculate:math:argb -12, Calculate:math:operator -12, Calculate:mode -12, Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/212/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-12
CALCulate:MATH:ARGA
Description:
Set or return the signal source of the A argument for use in channel math. Channel math operates
pixel by pixel on two source channels to produce a third channel result
Syntax:
CALCulate:MATH:ARGA
Argument:
Valid Modes:
Any
CALCulate:MATH:ARGB
Description:
Set or return the signal source of the B argument for use in channel math. Channel math operates
pixel by pixel on two source channels to produce a third channel result
Syntax:
CALCulate:MATH:ARGB
Argument:
Valid Modes:
Any
CALCulate:MATH:OPERator
Description:
Set or return the pixel-by-pixel mathematical operation performed between sources A and B to
create the math channel result. Note that the apparent operation performed is different for
logarithmic and non-logarithmic units for equivalent results.
Syntax:
CALCulate:MATH:OPERator
Argument:
Log units
Non-log units
CH_SUM
------
add
CH_DIFF ------
subtract
CH_RAT subtract
divide
Valid Modes:
Any
CALCulate:MODE
Description:
Set or return the system measurement mode. MODULATED is a continuous measurement mode,
PULSE is a triggered, oscilloscope-like mode, and STATISTICAL, which includes CDF, CCDF
and PDF as sub-modes, may be operated in a start-stop continuous mode, a decimated continuous
mode or a gated (triggered) mode.
Syntax:
CALCulate:MODE
Argument:
Note: The STATISTICAL sub-mode CDF, CCDF or PDF is restored but not reported.
Valid Modes:
Any