Table 4-8b. statistics > menu, Table 4-8b. statistics > menu -39, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual
Page 111: Operation
Boonton 4500B RF Peak Power Analyzer
Operation
4-39
Table 4-8B. Statistics > Menu
Menu Item
Function
(Type)
Selections
(SCPI cmd)
Presentation
Stat
Linear or logarithmic x-axis probability display
(Toggle)
Lin, Log
Select Lin to display percent probability on a linear scale. Select Log to
display percent probability on a logarithmic scale.
DISPlay:STATistical:PRESentation {LIN, LOG}
Horiz Scale
Stat
Select horizontal scale
(Numeric)
Lin: Discrete range
0.1% / Div to 10% / Div Use any of the data entry controls to select a horizontal range. The
10% / Div range displays the total scope of probability values from
0% to 100%. For the more sensitive ranges, also use the % Offset
function (see, below) to position the expanded scale so that the point
of interest is on-screen.
An invalid combination of Horiz Scale and Horiz Offset is not allowed.
The Offset will be forced to a valid value.
DISPlay:PCTSCALE <10, 5, 2, 1, 0.5, 0.2, 0.1> (%)
Log: Discrete range
Use any of the data entry controls to select the number of decades of
1 to 7 decades
probability displayed on the horizontal axis.
DISPlay:DECAdes <1 to 7> (decades)
Full Scale
Stat
Select the right edge Horiz Scale (Log) value
(Numeric)
Discrete decade ranges:
Log: 0.01 % to 100 %
For the logarithmic scale only, use any of the data entry controls to
select the percent value of the right hand edge of the display. Allows
the display to be shifted to the right by 4 additional decades to display
probability down to 1E-9 %. See Figure 4-25b.
DISPlay:FULLscale <.01 to 100> (%)