Display:header, Display:help, Display:mode – Boonton 4500B Peak Power Meter User Manual
Page 220: Display:modulated:timebase, Display:header -20, Display:help -20, Display:mode -20, Display:modulated:timebase -20, Boonton 4500b rf peak power analyzer, Remote operation
![background image](/manuals/401804/220/background.png)
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-20
DISPlay:HEADer
Description:
Set or return the type of header shown at the top left of the display.
Syntax:
DISPlay:HEADer
Argument:
internal temperature)}
Valid Modes:
All
DISPlay:HELP
Description:
Set the display to the show the Help page instead of Graphics or Text page. Help provides a short
description of each menu item. Set ON to see the HELP page; set OFF to restore the previous
GRAPH or TEXT page.
Syntax:
DISPlay:HELP
Argument:
Valid Modes:
All
DISPlay:MODE
Description:
Set or return the Text or Graphics page display setting. The Graphics page displays traces and
Marker measurements. The Text page displays auto-measure values in text form.
Syntax:
DISPlay:MODE
Argument:
Valid Modes:
All
DISPlay:MODUlated:TIMEBASE
Description:
Set or return the modulated mode timebase in seconds/division. The 4500B has fixed timebase
settings in a 1-2-5 sequence, and if the argument does not match one of these settings, it will be
forced to the next highest entry.
Syntax:
DISPlay:MODUlated:TIMEBASE
Argument:
30 sec/div,
1 to 10 min/div,
30 min/div and 60 min/div
Valid Modes:
Modulated mode
Note:
There are separate time bases for the Modulated mode and the Pulse/Gated Statistical
modes. The arguments selected are saved and restored independently by mode.