Display:trace:pdfbars, Display:trace:type, Display:trace:vcenter – Boonton 4500B Peak Power Meter User Manual
Page 231: Display:trace:vscale, Display:trace:pdfbars -31, Display:trace:type -31, Display:trace:vcenter -31, Display:trace:vscale -31, Boonton 4500b rf peak power analyzer, Remote operation
Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-31
DISPlay:TRACe:PDFBars
Description:
Set or return the number of bars in the PDF display.
Syntax:
DISPlay:TRACe[1|2]:PDFBars
Argument:
Valid Modes:
Statistical PDF mode only.
DISPlay:TRACe:TYPE
Description:
Set or return the trace drawing style, points, lines or off. For Points, display pixels are plotted
individually. For Lines, display pixels are connected with vertical lines. For Off pixels are not
plotted on the display, however, measurements continue to be made.
Syntax:
DISPlay:TRACe[1|2|3|4|5|6|7]:TYPE
Argument:
Valid Modes:
All
DISPlay:TRACe:VCENTer
Description:
Set or return the power or voltage level of the horizontal centerline of the graph for the specified
channel in channel units. If a change in the vertical scale causes the center maximum value to be
exceeded, the center will be forced to the maximum value for the new range.
Syntax:
DISPlay:TRACe[1|2|3|4|5|6|7]:VCENTer
Argument:
10,000 times the vertical scale power/div for watts units
10,000 times the vertical scale volts/div for volts units
Valid Modes:
All
DISPlay:TRACe:VSCALe
Description:
Set or return the power or voltage vertical sensitivity of the trace display in channel units.
Syntax:
DISPlay:TRACe[1|2|3|4|5|6|7]:VSCALe
Argument:
Units = dBm, dBV, dBmV, dBuV ,
Units = watts
Units = volts
Valid Modes:
All