beautypg.com

Table 4-29. util > hardcopy > submenu, Table 4-29. util > hardcopy > submenu -81, Boonton 4500b rf peak power analyzer – Boonton 4500B Peak Power Meter User Manual

Page 153: Operation

background image

Boonton 4500B RF Peak Power Analyzer

Operation

4-81

Table 4-29. Util > Hardcopy > Submenu

Menu Item

Function

(Type)

Selections

(SCPI cmd)


Plot Labels

Show optional Plot Labels on the display

(Toggle)

On, Off

and the hardcopy

DISPlay:PLABel:MODE {ON, OFF}




Plot BackGrnd

Select the background color for hardcopy

(Mult. Choice)

Black, White, Gray

For printing and plotting, the black display

background may be undesirable. Use this

menu to select a black, gray or white back-
ground for the hardcopy.


HCOPy:BACKground:COLor
{BLACK, WHITE, GRAY}















Fig. 4-43. Util >
Hardcopy
Submenu


Related commands to enter up to 4 custom labels:

DISPlay:PLABel:TEXTA ―string of up to 19 characters‖
DISPlay:PLABel:TEXTB ―string of up to 19 characters‖
DISPlay:PLABel:TEXTC ―string of up to 19 characters‖
DISPlay:PLABel:TEXTD ―string of up to 19 characters‖