Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual
Page 103

Boonton 4500B RF Peak Power Analyzer
Operation
4-31
Table 4-7. Chan Mem # > Menu (continued)
Menu Item
Function
(Type)
Selections
(SCPI cmd)
Waveform
Pulse, Mod & Stat
Displays a report that describes the instrument setup for the
(Action)
REPORT
waveform stored in Mem #.
This text report typically includes the waveform source and type, the
instrument serial number, the sensor model number and serial number,
channel, dB Offset, Frequency, Frequency CF, video bandwidth,
averaging, timebase, position, trigger delay, trigger mode, trigger
source, trigger level, trigger holdoff, and trigger slope. For statistical
distributions trigger information is replaced by running time
.
and number of samples. Format details vary with mode. See Fig 4-19.
Extensions
Pulse, Mod & Stat
Accesses the Chan Mem # > Extensions > submenu. See Table 4-7A.
(Submenu)
Table 4-7A. Chan Mem # > Extensions > Submenu
Menu Item
(Type)
Selections
Function
Units
Pulse, Mod & Stat
Identical to Chan # > Exten > Units menu. See Table 4-4.
The stored waveform may be displayed in any of the units allowed for
the original waveform.
Define Pulse
Pulse & Mod
Identical to Chan # > Exten > Define Pulse menu. See Table 4-4.