Boonton 4500b rf peak power analyzer, Operation – Boonton 4500B Peak Power Meter User Manual
Page 190
Boonton 4500B RF Peak Power Analyzer
Operation
4-118
AUTOMATIC MEASUREMENTS
– STATISTICAL MODE (Opt 10 only)
In the Statistical mode automatic measurements are made only in the ―Run‖ mode. When the Top Level > Measurement >
Stop selection is made, the latest value is displayed. Press the Top Level > Display > CLEAR menu key to clear previous
results. Press the Top Level > Measurement > Stop menu key to select ―Run‖. The measurement will begin and continue
until a terminate condition is met. The automatic measurements will continuously update with the results of the accumulating
data. The peak-to-average reading is not measured on External Trigger channels. The peak-to-average and dynamic range
results are always displayed in dB units. The average, peak and minimum results will have the same units as the associated
channel.
Table 4-61. Statistical Mode
– Automatic Measurements I
CH1&2
CH 1&2
Name
(Power Units)
(Voltage Units)
Avg
Average power
Average voltage (converted)
Peak
Peak power
Peak voltage
Min
Minimum power
Minimum voltage
Pk/Avg
Peak-to-average ratio (dB)
Peak-to-average ratio (dB)
Dyn Rng
Dynamic range (dB)
Dynamic range (dB)
Table 4-62. Statistical Mode
– Automatic Measurements II
EXT TRIG 1&2
Name
(Voltage Units)
Avg
Average voltage
Peak
Peak voltage
Min
Minimum voltage
Pk/Avg
----
Dyn Rng
Peak-to-peak voltage