Power:l2, Power:set – GW Instek PEL-3000 Series Programming Manual User Manual

Page 95

background image

COMMNAND OVERVIEW

93

Example

:POWer:L1 MIN
Sets the power value to the minimum.

Query example

:POWer:L1?
>10W
Returns the “Lewel1” power setting.

:POWer:L2

Set

Query

Description

Sets or queries the CP mode “Level2” power. This
command is only applicable to dynamic mode.
Note: A different “Level2” value can be set to
different current ranges.

Syntax

:POWer:L2 {[ W ] | MINimum | MAXimum }

Query Syntax

:POWer:L2?

Parameter

[ W ]

“Level2” power value

MINIMUM or MIN

Minimum power level

MAXIMUM or MAX

Maximum power level

Return parameter

Returns the “Level2”
power value.

Example

:POWer:L2 MIN
Sets the power value to the minimum.

Query example

:POWer:L2?
>10W
Returns the “Level2” power setting.

:POWer:SET

Set

Query

Description

Sets or queries the CP dynamic mode Set value
when the dynamic units are set to Percent.
This command is only applicable in dynamic
mode.

Syntax

:POWer:SET {[ W ] | MINimum | MAXimum }

Query Syntax

:POWer:SET?