COMMNAND OVERVIEW

43

Example1

:OPP LIMIT
Sets the OCP setting to limit.

Example2

:OPP 10.000
Sets the OPP value to 10W.

Query example

:OPP?
>LIMIT, 10.000
The OPP setting is LIMIT and the OPP value is
10.000W.

[:CONFigure]:UVP

Set

Query

Description

Sets or queries the UVP trip settings. The UVP can
also be cleared with this command.

Syntax

[:CONFigure]:UVP {[ V ] |MINimum |
MAXimum }

Query Syntax

[:CONFigure]:UVP?

Parameter

[V]

voltage limit value.

MINIMUM or MIN

MINIMUM value.

MAXIMUM or MAX

MAXIMUM value.

Return parameter Returns the UVP level ()
Example1

:UVP 10.00
Sets the UVP setting to 10V.

Query example

:UVP?
> 10.0000
The UVP setting is 10.0000V.

[:CONFigure]:OVP

Set

Query

Description

Sets or queries the OVP trip settings. The OVP can
also be cleared with this command.

Syntax

[:CONFigure]:OVP {[ V ] | MINimum|
MAXimum}