COMMNAND OVERVIEW

91

Power Subsystem Commands

:POWer[:VA] ......................................................................... 91

:POWer:VB............................................................................ 92
:POWer:L1 ............................................................................. 92

:POWer:L2 ............................................................................. 93
:POWer:SET ......................................................................... 93

:POWer:LEVel ...................................................................... 94
:POWer:T1 ............................................................................ 95

:POWer:T2 ............................................................................ 95
:POWer:FREQuency ........................................................... 96

:POWer:DUTY ..................................................................... 97

:POWer[:VA]

Set

Query

Description

Sets or queries the CP mode “A Value” power.
This command is applicable to static mode only.
Note: A different “A Value” can be set for different
current ranges.

Note: The [:VA] node can only be omitted when in
static mode.

Syntax

:POWer[:VA] {[ W ] | MINimum | MAXimum }

Query Syntax

:POWer[:VA]?

Parameter

[ W ]

“A Value” power value

MINIMUM or MIN

Minimum power level

MAXIMUM or MAX

Maximum power level

Return parameter

Returns the “A Value”
power value.

Example

:POWer:VA MIN
Sets the power value to the minimum.

Query example

:POWer:VA?
>10W
Returns the “A Value” power setting.