PEL-3000 Programming Manual

70

Current Subsystem Commands

:CURRent[:VA] ..................................................................... 70

:CURRent:VB ........................................................................ 71
:CURRent:SRATe ................................................................. 71

:CURRent:L1 ......................................................................... 72
:CURRent:L2 ......................................................................... 72

:CURRent:SET ...................................................................... 73
:CURRent:LEVel .................................................................. 74

:CURRent:RISE .................................................................... 74
:CURRent:FALL ................................................................... 75

:CURRent:T1 ......................................................................... 75
:CURRent:T2 ......................................................................... 76

:CURRent:FREQuency ........................................................ 77
:CURRent:DUTY ................................................................. 77

:CURRent[:VA]

Set

Query

Description

Sets or queries the CC mode “A Value” (static
mode) current. This command is applicable to
static modes.
Note: A different current value can be set for each
current range (High/Mid/Low).

Note: The [:VA] node can only be omitted when in
static mode.

Syntax

:CURRent[:VA] {[ A ] | MINimum | MAXimum }

Query Syntax

:CURRent[:VA]?

Parameter

[A]

“A Value” current value

MINIMUM or MIN

Minimum current level

MAXIMUM or MAX

Maximum current level

Return parameter

Returns the “A Value”
current value.

Example

:CURRent:VA MIN
Sets the current value to the minimum.