PEL-3000 Programming Manual

58

Parameter

[ A ]

Step resolution.

MINIMUM or MIN

Minimum step resolution

MAXIMUM or MAX

Maximum step resolution

Return parameter

Returns the range and the
step resolution.

Example

:STEP:CCL 0.02mA
Sets the step resolution to 0.02mA.

Query example

:STEP:CCL?
> CCL:0.00002
Returns the step resolution (0.00002A).

[:CONFigure]:STEP:CC

Set

Query

Description

Returns the step resolution for each CC Mode
range as a string.

Query Syntax

[:CONFigure]:STEP:CC?

Return parameter

Returns the CCH, CCM
and CCL step resolution
settings.

Query example

:STEP:CC?
>CCH:0.002, CCM:0.0002, CCL:0.00002
Returns the CC mode step resolution for each range.

[:CONFigure]:STEP:CRH

Set

Query

Description

Configures the step resolution for CR High Range.
Note: The step resolution setting will be
automatically rounded to the closest multiple of
the base resolution.

Syntax

[:CONFigure]:STEP:CRH {[ MHO ] | MINimum
| MAXimum }

Query Syntax

[:CONFigure]:STEP:CRH?

Parameter

[ MHO ]

Step resolution. [MHO]

MINIMUM or MIN

Minimum step resolution

MAXIMUM or MAX

Maximum step resolution