beautypg.com

Resistance:duty – GW Instek PEL-3000 Series Programming Manual User Manual

Page 89

background image

COMMNAND OVERVIEW

87

Query Syntax

:RESistance:FREQuency?

Related
Commands

:RESistance:DUTY

Parameter

Sets the switching
frequency in hertz.

MINIMUM or MIN

Minimum frequency

MAXIMUM or MAX

Maximum frequency

Return parameter

Returns the frequency in
hertz.

Example

:RESistance:FREQuency 60
Sets frequency to 60Hz.

Query example

:RESistance:FREQuency?
>60
Returns the switching frequency (60Hz).

:RESistance:DUTY

Set

Query

Description

Sets the positive duty cycle for the CR dynamic
mode switching frequency. This command along
with the “:RESistance:FREQuency” command is
used when the dynamic time setting is set to
“Freq./Duty”.
This command is only applicable for dynamic
mode.

Syntax

:RESistance:DUTY { | MINimum | MAXimum}

Query Syntax

:RESistance:DUTY?

Related
Commands

:RESistance:FREQuency

Parameter

Sets the duty as a
percentage.

MINIMUM or MIN

Minimum duty

MAXIMUM or MAX

Maximum duty

Return parameter

Returns the duty as a
percentage.