Current:vb, Current:srate – GW Instek PEL-3000 Series Programming Manual User Manual

Page 73

background image

COMMNAND OVERVIEW

71

Query example

:CURRent:VA?
>1.0A
Returns the “A Value” current setting.

:CURRent:VB

Set

Query

Description

Sets or queries the CC mode “B Value” (static
mode) current. This command is applicable to
static modes.
Note: A different current value can be set for each
current range (High/Mid/Low).

Syntax

:CURRent:VB {[ A ] | MINimum | MAXimum }

Query Syntax

:CURRent:VB?

Parameter

[A]

“B Value” current value

MINIMUM or MIN

Minimum current level

MAXIMUM or MAX

Maximum current level

Return parameter

Returns the “B Value”
current value.

Example

:CURRent:VB MIN
Sets the current value to the minimum.

Query example

:CURRent:VB?
>1.0A
Returns the “B Value” current setting.

:CURRent:SRATe

Set

Query

Description

Sets or queries the current slew rate for CC static
mode.

Syntax

:CURRent:SRATe {| MINimum | MAXimum }

Query Syntax

:CURRent:SRATe?

Parameter

Sets the slew rate in
mA/us

MINIMUM or MIN

Set to the highest slew rate

MAXIMUM or MAX

Set to the lowest slew rate