PEL-3000 Programming Manual

30

*SAV

Set

Description

The Save Instrument State command saves the
instrument settings to one of the memory setting
slots.

Syntax

*SAV

Parameter

1~256

Memory number 1 to 256

Example

*SAV 20
Saves the current setting to memory 20.

*SRE

Set

Query

Description

Queries or sets the Service Request Enable register.
The Service Request Enable register determines
which events in the Status Byte register can set the
Master Summary bit (MSB) in the Status Byte
Register. Any bits that are set to 1 will cause the
MSS bit to be set.

Syntax

*SRE

Query Syntax

*SRE?

Parameter/
Return parameter

(bit weight)

Bit number / Description

1

0/Not used

2

1/ERR

4

2/ CSUM

8

3/ QUES

16

4/ MAV

32

5/ ESB

64

6/ RQS_MSS

128

7/ OPER

Note: Bit 1 and 2 cannot be set, however bit 1 (ERR)
can be returned.

Example

*SRE 8
Sets bit 3 of the Service Request Enable register.