PEL-3000 Programming Manual

52

Query example

:GNG?
>GO
Returns the Go-NoGo test result.

[:CONFigure]:GNG:H

Set

Query

Description

Sets or queries the high voltage/current limit
value. If the entry mode is set to value, the high
voltage/current limit value units are in
volts/amps. If the entry mode is set to percent, the
high voltage/current limit value units are in
percent.

Syntax

[:CONFigure]:GNG:H

Query Syntax

[:CONFigure]:GNG:H?

Parameter

Sets the high
voltage/current limit value
in volts/amps or in
percent.

Return parameter

Returns the high
voltage/current limit value
in volts/amps or as %.

Example

:GNG:H 100.0
Sets the high voltage limit value to 100%.

Query example

:GNG:H?
>100.0
Returns the high voltage limit value as 100.0%.

[:CONFigure]:GNG:L

Set

Query

Description

Sets or queries the low voltage/current limit value.
If the entry mode is set to value, the low
voltage/current limit value units are in
volts/amps. If the entry mode is set to percent, the
low voltage/current limit value units are in
percent.

Syntax

[:CONFigure]:GNG:L