beautypg.com

Voltage subsystem commands, Voltage[:va, Voltage:vb – GW Instek PEL-3000 Series Programming Manual User Manual

Page 91

background image

COMMNAND OVERVIEW

89

Voltage Subsystem Commands

:VOLTage[:VA] ..................................................................... 89

:VOLTage:VB ....................................................................... 89

:VOLTage[:VA]

Set

Query

Description

Sets or queries the CV mode “A Value” voltage or
the +CV voltage value. Note: The same value
applies for each current range (High/Mid/Low).

Note: The optional command node [:VA] can only be
omitted when in static mode.

Syntax

:VOLTage[:VA] {[ V ] | MINimum | MAXimum }

Query Syntax

:VOLTage[:VA]?

Parameter

[ V ]

“A Value” voltage value

MINIMUM or MIN

Minimum voltage level

MAXIMUM or MAX

Maximum voltage level

Return parameter

Returns the “A Value”
voltage value.

Example

:VOLTage:VA MIN
Sets the voltage value to the minimum.

Query example

:VOLTage:VA?
>1.000V
Returns the “A Value” voltage setting.

:VOLTage:VB

Set

Query

Description

Sets or queries the CV mode “B Value”. Note: The
same value applies for each current range
(High/Mid/Low).

Syntax

:VOLTage:VB {[ V ] | MINimum | MAXimum }

Query Syntax

:VOLTage:VB?