beautypg.com

Trigger commands, Trigger states, Abort – GW Instek PEL-3000 Series Programming Manual User Manual

Page 35

background image

COMMNAND OVERVIEW

33

Trigger Commands

:ABORt .................................................................................. 33

:INPut[:STATe]:TRIGgered ............................................... 34
:INITiate[:IMMediate] .......................................................... 34

:INITiate:CONTinuous ....................................................... 34
:CURRent[:VA]:TRIGgered ................................................ 35

:RESistance[:VA]:TRIGgered ............................................. 35

Trigger States

Idle state

Trigger Wait

state

:INIT:CONT

PEL-3000

power on

Triggered

ON

OFF

*RST,

ABORt

:INITate[:IMMediate],

:INITiate:CONTinuous ON

*TRG,

manual trigger

:ABORt

Set

Description

Clears the trigger wait status and returns to the
idle state.

Query Syntax

:ABORt