beautypg.com

Configure]:gng:mode, Configure]:gng[:pass – GW Instek PEL-3000 Series Programming Manual User Manual

Page 53

background image

COMMNAND OVERVIEW

51

Return parameter

Returns the delay time in
seconds.

Example

:GNG:DTIMe 0.5
Sets the delay time to 0.5 seconds.

Query example

:GNG:DTIMe?
>0.5
The delay time is 0.5 seconds.

[:CONFigure]:GNG:MODE

Set

Query

Description

Sets or queries the entry mode for the Go-NoGo
settings. The entry mode determines whether the
Go-NoGo limits are set as values or as a
percentage value from a center reference value.

Syntax

[:CONFigure]:GNG:MODE {PERCent | VALue }

Query Syntax

[:CONFigure]:GNG:MODE?

Parameter

PERCENT or PERC

Sets the entry mode to %.

VALUE or VAL

Sets the entry mode to
value

Return parameter Percent

% entry mode

Value

Value entry mode

Example

:GNG:MODE PERCent
Sets the entry mode to %.

Query example

:GNG:MODE?
>Percent
The entry mode is %.

[:CONFigure]:GNG[:PASS]

Set

Query

Description

Queries the Go-NoGo test result(s). This command
can be used for all test modes (CC, CV, CR, CP).

Query Syntax

[:CONFigure]:GNG[:PASS]?

Return parameter NG

No Go (fail)

GO

Go (Pass)