COMMNAND OVERVIEW

65

External Control Commands

[:CONFigure]:EXTernal[:CONTrol] ................................. 65

[:CONFigure]:EXTernal:LOADonin ................................ 65

[:CONFigure]:EXTernal[:CONTrol]

Set

Query

Description

Configures the unit for external control or disables
external control.

Syntax

[:CONFigure]:EXTernal[:CONTrol] {OFF | VOLTage |
RESistance | RINV }

Related
Commands

[:CONFigure]:EXTernal[:CONTrol]?

Parameter

OFF

Disables external control

VOLTAGE or VOLT

Sets the unit to external
voltage control

RESISTANCE or RES

Sets the unit to external
resistance control

RINV

Sets the unit to external
resistance (inverted)
control

Return Parameter

Returns the external
control configuration

Example

:EXTernal VOLT
Turns external control voltage control on.

Query example

:EXTernal?
>voltage
The unit uses external voltage control.

[:CONFigure]:EXTernal:LOADonin

Set

Query

Description

The LOADonin settings determine whether the
load is turned on when an external switch is closed
(LOW) or open (HIGH).