beautypg.com

Load:protection:clear, Load:time – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 77

background image

COMMAND DETAILS

77

Description

Returns the protection levels for electronic load

Query Syntax

:LOAD:PROTection?

Query Example

:LOAD:PROT?

Clears the Channel
Status Register.

Return Parameter

Condition

Condition

1

OC

16

OT

2

OV

32

G/N

4

OP

64

UVP

8

RV

16-128

Not Used

Query Example

:LOAD:PROT?
0

Returns the status of the
Channel Status Register.
Here 0 is returned
indicating no protection
settings have been
tripped.

:LOAD:PROTection:CLEar

Channel Specific
Command

Description

This command clears the Channel Status Register
for the specific channel. See page162.

Syntax

:LOAD:PROTection:CLEar

Example

:LOAD:PROT:CLE

Clears the Channel
Status Register.

:LOAD:TIME?

Channel Specific
Command

Description

This command displays the total load on time. If
the load is on, the load time when the command
was issued is displayed.

Query Syntax

:LOAD:TIME?

Return Parameter 1unit = 1 second

Load on time

2.2

2.2 seconds