Resistance[:static]:rise/fall – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 104

background image

PEL-2000 Programming Manual

104

Example

:RES:L1 10

Sets CR static mode A
Value to 10 ohms,
depending on the
specific range

Query Syntax

:RESistance[:STATic]:L1/L2? [MAX|MIN]

Return Parameter [MAX|MIN]

Resistance

1 unit = 1 ohm/1 k ohm

Returns the resistance of
the A or B Value.

MAX/MIN

Returns the allowable
maximum and minimum.

Query Example

:RES:L1? MAX
300

Returns the maximum
resistance allowed for
the channel. (PEL-2020)

:RESistance[:STATic]:RISE/FALL

Channel Specific
Command

Description

Sets the rising/falling slew rate for CR mode. The
command applies to the current mode
(static/dynamic) and the current range
(High/Low)

Syntax

:RESistance[:STATic]:RISE/FALL [A/uS]

Parameter

[A/uS]

Slew rate

RISE/FALL 0.8A/uS

Sets the rising/falling slew
rate to 0.8A/uS

RISE/FALL .8

Sets the rising/falling slew
rate to 0.8A/uS

RISE/FALL MIN

Sets to the slowest
rising/falling slew rate.

RISE/FALL MAX

Sets to the fastest
rising/falling slew rate.

Example

:RES:RISE 0.1

Sets the rising slew rate
to 0.1A/uS.

Query Syntax

:RESistance:RISE/FALL? [MIN|MAX]