beautypg.com

Load:short:key, Load:protection – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 76

background image

PEL-2000 Programming Manual

76

Parameter

ON/1

Shorting is On

OFF/0

Shorting is Off

Example

:LOAD:SHOR ON

Short circuits the load
module channel.

Query Syntax

:LOAD:SHORt[:STATe]?

Return Parameter

Short Load module

1

Shorting is activated

0

Shorting is deactivated

Query Example

:LOAD:SHOR?
0

Shorting is deactivated
on the specific channel.

:LOAD:SHORt:KEY

Channel Specific
Command

Description

The SHORT key can be set to Toggle or Hold
mode.

Syntax

:LOAD:SHORt:KEY {TOGGLE|1|HOLD|0}

Parameter

TOGGLE/1

Sets the SHORT key to
toggle mode

HOLD/0

Sets the SHORT key to
hold mode

Example

:LOAD:SHOR:KEY 1

Set the SHORT key to
toggle.

Query Syntax

:LOAD:SHORt:KEY?

Return Parameter

Mode

1

Toggle mode is active

0

Hold mode is active

Query Example

:LOAD:SHOR:KEY?
0

Hold mode is active for
the specific channel.

:LOAD:PROTection?

Channel Specific
Command