beautypg.com

Resistance:static:recall, Resistance:static:low:avalue/bvalue – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 105

background image

COMMAND DETAILS

105

Return Parameter [MAX|MIN]

Slew rate

1 Unit =1 amp/uS

Returns the slew rate.

MAX/MIN

Returns the allowable
maximum and minimum.

Query Example

:RES:RISE? MAX
0.8000

Returns the maximum
value for the rising slew
rate (0.8 A/uS).

:RESistance:STATic:RECall

Channel Specific
Command

Description

Sets or queries whether A Value or B Value is the
currently active value in CR static mode.

Syntax

:RESistance:STATic:RECall {A|0|B|1}

Parameter

A/0

A

B/1

B

Example

:RES:STAT:REC 1

Makes B Value the active
value.

Query Syntax

:RES:STATic:RECall?

Return Parameter

Value

0

A

1

B

:RESistance:STATic:LOW:AVALue/BVALue

Channel Specific
Command

Description

Sets the low range A/B Value for constant
resistance static mode.

Syntax

:RESistance:STATic:LOW:AVALue/BVALue
[OHM]

Parameter

NRf+[OHM]

Resistance

AVALue 10

Sets A Value to 10 ohms.
(Low range only)