PEL-2000 Programming Manual

62

RISE/FALL MIN

Sets to the slowest
rising/falling slew rate.

RISE/FALL MAX

Sets to the fastest
rising/falling slew rate.

Example

:CURR:STAT:HIGH:RISE
1.1

Sets the rising slew rate
to 1.1A/uS.

Query Syntax

:CURRent:STATic:HIGH:RISE/FALL? [MIN|MAX]

Return Parameter [MAX|MIN]

Slew rate

1 Unit=1 amp/uS

Returns the slew rate.

MAX/MIN

Returns the allowable
maximum and minimum.

Query Example

:CURR:STAT:HIGH:RISE?
MAX
0.8000

For high range CC mode,
the maximum value for
the rising slew rate is
0.8000 A/uS for the
specific channel.

:CURRent:DYNamic:L1/L2

Channel Specific
Command

Description

Sets the current levels (Level 1 & 2) for CC
dynamic mode. The command is range dependant.
If the current range is Low, then the settings will
only apply to low range.

Syntax

:CURRent:DYNamic:L1/L2 [A]

Parameter

NRf+[A]

Current

L1 1

Sets L1 to 1 Amp.

L2 2

Sets L2 to 2 Amps.

L2 2A

Sets L2 to 2 Amps.

L1/L2 MIN

Sets L1 or L2 to the
minimum level for the
specific channel.