COMMAND DETAILS

135

Power Subsystem

:POWer:L1/L2 ............................................................... 135

:POWer:CURRent ......................................................... 136

:POWer:RECall .............................................................. 137

:POWer:LOW:AVALue/BVALue .................................. 137

:POWer:LOW:CURRent ............................................... 138

:POWer:HIGH:AVALue/BVALue ................................ 139

:POWer:HIGH:CURRent .............................................. 140

:POWer:L1/L2

Channel Specific
Command

Description

Sets the A/B Value for constant power mode,
where L1 is A Value and L2 is B Value. The
command is range dependant. If the current range
is Low, then the command will only apply to the
low range settings.

Syntax

:POWer:L1|L2 [W]

Parameter

[W]

L1 1

Sets A Value to 1 Watt.

L2 2

Sets B Value to 2 Watts.

L1 1W

Sets A Value to 1 Watt.
(single channel only)

L1 MIN

Sets A Value to the
minimum level for the
specific channel.

L1 MAX

Sets A Value to the
maximum Level for the
specific channel.

Example

:POW:L1 1

Sets A Value to 1 watt

Query Syntax

:POW:L1?/L2? [MAX|MIN]

Return Parameter [MAX|MIN]

Current