Sequence:end, Sequence:point:resistance – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 150

background image

PEL-2000 Programming Manual

150

Query Example

:SEQ:EDIT:POIN?
3

The current point is 3.

:SEQuence:END

Channel Specific
Command

Description

Sets the number of points in the sequence.

Syntax

:SEQuence:END

Parameter

Points

1~120

1~120.

Example

:SEQ:END 5

Sets the max number of
points to 5.

Query Syntax

:SEQuence:END?

Return Parameter

Points

1~120

1~120

Query Example

:SEQ:END?
5

There are 5 points in the
sequence.

:SEQuence:POINt:RESistance

Channel Specific
Command

Description

Sets the resistance value for the current point. CR
mode only.

Syntax

:SEQuence:POINt:RESistance
[OHM]|MIN|MAX

Parameter

[OHM], MIN, MAX Resistance value

100

100Ω

100 OHM

100Ω

MAX/MIN

Maximum or minimum
value.

Example

:SEQ:POIN:RES 100

Sets the resistance to 100.

Query Syntax

:SEQuence:POINt::RESistance? [MAX|MIN]

Return Parameter

Resistance Value