beautypg.com

Resistance:dynamic:high:l1/l2 – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 111

background image

COMMAND DETAILS

111

Parameter

[S]

Time T1/T2

T1/T2 0.1S

Sets the T1/T2 time to 0.1
seconds.

T1/T2 1

Sets T1/T2 to 1 second.

T1/T2 MIN

Sets the T1/T2 to the
minimum value.

T1/T2 MAX

Sets the T1/T2 time to the
maximum time

Example

:RES:DYNA:LOW:T1 10S

Sets the T1 time to 10
seconds for the specific
channel.

Query Syntax

:RESistance:DYNamic:T1/T2? [MIN|MAX]

Return Parameter [MAX|MIN]

Time T1/T2

1 Unit=1 second

Returns T1/T2 time.

MAX/MIN

Returns the allowable
maximum and minimum.

Query Example

:RES:DYN:LOW:T1?
2.5
:CURR:DYN:LOW:T1?
MIN
0.000025

Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.

:RESistance:DYNamic:HIGH:L1/L2

Channel Specific
Command

Description

Sets the high range resistance levels (Level 1 & 2)
for CR dynamic mode.

Syntax

:RESistance:DYNamic:HIGH:L1/L2 [OHM]

Parameter

NRf+[OHM]

Resistance

L1 10

Sets L1 to 10 ohms. (high
range only)

L2 20OHM

Sets L2 to 20 ohms. (high
range only)