beautypg.com

Current:dynamic:rise/fall – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 63

background image

COMMAND DETAILS

63

L1/L2 MAX

Sets L1 or L2 to the
maximum Level for the
specific channel.

Example

:CURR:DYN:L1 10

In CC dynamic mode,
Set L1 (level 1) to 10
amps.

Query Syntax

:CURRent:DYNamic:L1/L2? [MIN|MAX]

Return Parameter [MAX|MIN]

Current

MAX/MIN

Returns the allowable
maximum and minimum.

1 unit= 1 amp

Returns the current of
L1/L2, or the maximum or
minimum current allowed.

Query Example

:CURR:DYN:L2?
2.0400

Returns current for the
specific channel.

:CURRent:DYNamic:RISE/FALL

Channel Specific
Command

Description

Sets the rising/falling slew rate for CC dynamic
mode for the specific channel and range.

Syntax

:CURRent:DYNamic:RISE/FALL [A/uS]

Parameter

[A/uS]

Slew Rate

RISE/FALL 0.8A/uS

Sets the rising/falling slew
rate to 0.8A/uS

RISE/FALL 1

Sets the rising/falling slew
rate to 1A/uS

RISE/FALL MIN

Sets to the slowest
rising/falling slew rate.

RISE/FALL MAX

Sets to the fastest
rising/falling slew rate.

Example

:CURR:DYNA:RISE 1.1

Sets the rising slew rate
to 1.1A/uS.

Query Syntax

: CURRent:DYNamic:RISE/FALL? [MIN|MAX]